Propaq CS Directions for Use
37
Monitoring
NIBP
If artifact is so severe while Smartcuf is enabled that it affects the accuracy of an NIBP measurement,
that measurement is marked with a special symbol on the display and on printouts:
There may be some situations where it is desirable to disable Smartcuf. This may include situations
with very extreme motion artifact, certain types of arrhythmias, or other situations where it is not
possible to obtain a good ECG signal. NIBP measurements can still be performed when Smartcuf is
disabled.
To disable Smartcuf, from the Main Menu press
NIBP
to display the NIBP Menu and set Smartcuf to
OFF.
Take the NIPB Reading
1.
Select a cuff and hose appropriate for the patient. Select cuff size based on limb circumference.
Use only hoses and cuffs listed in the Protocol Systems Products and Accessories booklet.
Note
Be sure the patient mode selected is appropriate for the cuff you are using. For instructions about changing the patient
mode, see page 22. Be aware that changing patient modes will cancel an NIBP reading in progress.
2.
Squeeze as much air from the cuff as you can before placing it on the patient.
Neonate Mode
Pediatric Mode
Adult Mode
Typical Hoses
Neonate/Infant
Neonate/Infant, Adult
Adult
Typical Cuffs
Neonate #1 to #5
(disposable); newborn,
infant (reusable)
Neonate #4, neonate #5,
infant, child, small adult
Child, small adult, adult,
large adult, thigh
Recommended Limb
Circumference
up to 15 cm
10 to 25 cm
greater than 15 cm
NIBP TREND
TIME
MON
10:01
9:46
9:31
9:20
9:08
8:52
8:47
8:41
8:35
8:30
HH:MM
HR/PR SpO
2
S D (M)
RR/BR
BPM
%
mmHg
Br/M
80
80
80
80
80
80
80
80
80
80
97
97
97
97
97
97
97
97
97
97
122 58 ( 85)
126 62 ( 86)
127 62 ( 86)
134 66 ( 89)
124 57 ( 86)
127 58 ( 88)
124 56 ( 85)
126 57 ( 84)
127 58 ( 85)
132 61 ( 89)
12
12
12
12
12
12
12
12
12
12
NEXT
TREND
This symbol indicates the NIBP
reading was taken in the
presence of high artifact while
monitoring ECG with Smartcuf
enabled. Artifact can affect
accuracy. To help reduce
artifact, see step 6 on page 39.
Trend Display
Printout
112
58
( 85)
NIBP
18:45
S
D
(M)
mmHg
Numeric Display
Summary of Contents for Propaq 242
Page 6: ...6 Protocol Systems Inc...
Page 18: ...18 Protocol Systems Inc...
Page 26: ...26 Protocol Systems Inc...
Page 50: ...50 Protocol Systems Inc...
Page 52: ...52 Protocol Systems Inc...
Page 64: ...64 Protocol Systems Inc...
Page 66: ...66 Protocol Systems Inc...
Page 72: ...72 Protocol Systems Inc...
Page 74: ...74 Protocol Systems Inc...
Page 78: ...78 Protocol Systems Inc...
Page 84: ...84 Protocol Systems Inc...
Page 86: ...86 Protocol Systems Inc...
Page 94: ...94 Protocol Systems Inc...
Page 120: ...120 Protocol Systems Inc...