48
Protocol Systems, Inc.
Set Up the CO
2
Display and Alarm Limits
Note
After you connect a Mainstream CO
2
sensor or Sidestream CO
2
watertrap, the Propaq CS monitor displays the
waveform briefly without a scale. It displays WARM UP (for Mainstream) or START UP (for Sidestream) in the CO
2
numerics
window. After about 30 seconds, the monitor displays the CO
2
measurement and waveform range.
CO
2
monitoring is typically displayed as shown:
1.
To adjust the display, from the Main Menu press
SpO2/CO2, CO2
to access the first CO
2
menu:
2.
Press
RANGE
to select the CO
2
waveform scale or range.
mmHg:
0-100
0-60 (default)
0-30
kPa:
0-14
0-8
0-4
%:
0-14
0-8
0-4
To change CO
2
units (mmHg, kPa, or %) see page 96.
Note
If an inspired value is displayed indicating patient rebreathing (non-zero INCO
2
), check the patient breathing circuit for
proper function. For Mainstream CO
2
, also remove the sensor from the patient’s airway, hold it away from any source of breath,
and confirm INCO
2
begins to go down to the baseline value. If the Propaq CS monitor continues to display inspired values,
return the Mainstream CO
2
sensor to Protocol Systems for service.
3.
Press
mm/s
to set the display sweep speed for CO
2
and RESP (3.13, 6.25, or 12.5 mm/sec). The
default is 6.25.
To view the sweep speed setting, press
MORE
to access the CO
2
status window.
NET
OFF
HR
MON
17:05:10
PRINT FAULT
II
BR
P
1
( 96)
BPM
mmHg
(M)
TEMP
100.4
98.6
1.8
97
121
79
S
D
P
2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
°
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
12
S
D
NIBP
ECG/
RESP
INVASIVE
PRESSURE
SpO
2
/
CO
2
SETUP
ADULT
ON NETWORK
T1
T2
∆
T
mV
cm
1
60
30
0
M
C
O
2
80
ALLEN KIM
MCO
2
mmHg
35
10
MCO2 indicates
Mainstream CO
2
is
active.
SCO2 is displayed
when Sidestream
CO
2
is active.
CO2 is displayed if a
Mainstream sensor
or Sidestream
watertrap is installed
but not active.
ETCO
2
INCO
2
appears here
when displayed. The
INCO
2
numeric value
is only displayed if it
is in alarm or if it is
≥
7.5 mmHg (or
1kPa or 1%).
Breath Rate is
determined from
the
CO
2
sensor.
Measured CO
2
levels are normally
displayed as a
waveform and an
ETCO
2
numeric
value
RANGE
mm/s
PREVIOUS
MENU
MORE
Summary of Contents for Propaq 242
Page 6: ...6 Protocol Systems Inc...
Page 18: ...18 Protocol Systems Inc...
Page 26: ...26 Protocol Systems Inc...
Page 50: ...50 Protocol Systems Inc...
Page 52: ...52 Protocol Systems Inc...
Page 64: ...64 Protocol Systems Inc...
Page 66: ...66 Protocol Systems Inc...
Page 72: ...72 Protocol Systems Inc...
Page 74: ...74 Protocol Systems Inc...
Page 78: ...78 Protocol Systems Inc...
Page 84: ...84 Protocol Systems Inc...
Page 86: ...86 Protocol Systems Inc...
Page 94: ...94 Protocol Systems Inc...
Page 120: ...120 Protocol Systems Inc...