Nurse Call Specifications
Nurse Call Cable Specifications
This cable (Part Number 008-0634-XX) must be customized by a biomedical technician to connect to
the local Nurse Call system.
Propaq CS Monitor Audible Alarm Indications
Alarm Condition
Tone
Patient alarm
Continuous ON
Apnea alarm
ON for 1 second, OFF for 1 second
Equipment alert
ON for 1 second, OFF for 4 seconds
Characteristic
Specification
Maximum switch current
1 A
Maximum switch voltage
30 V ac/dc
Isolation
1500 Vrms
Alarm relay
Energized during apnea alarm or patient alarm
1
1. Pressing the Suspend/Resume Alarm key or SUSPEND suspends the Nurse Call alarm for 90 seconds.
Customized cable
2
(Protocol Systems Part Number
008-0634-XX); see below.
2. Refer to the Protocol Systems Products and Accessories booklet to order the cable.
One end is a 4-pin plug compatible with the monitor
Nurse Call connector; the other end must be
customized to connect to the local Nurse Call system.
Nurse Call Connector
on Monitor
1 Normally Open
4 Not Connected
3 Normally Closed
2 Common (Arm)
30V at 1A MAX
Pin 4
Pin 3
Pin 1
Pin 2
EXTERNAL VIEW
1
2
3
4
Normally Open (Black)
Arm (Red)
Normally Closed (Green)
Not Connected
Nurse Call Cable (Part Number 008-0634-00)
Summary of Contents for Propaq 242
Page 6: ...6 Protocol Systems Inc...
Page 18: ...18 Protocol Systems Inc...
Page 26: ...26 Protocol Systems Inc...
Page 50: ...50 Protocol Systems Inc...
Page 52: ...52 Protocol Systems Inc...
Page 64: ...64 Protocol Systems Inc...
Page 66: ...66 Protocol Systems Inc...
Page 72: ...72 Protocol Systems Inc...
Page 74: ...74 Protocol Systems Inc...
Page 78: ...78 Protocol Systems Inc...
Page 84: ...84 Protocol Systems Inc...
Page 86: ...86 Protocol Systems Inc...
Page 94: ...94 Protocol Systems Inc...
Page 120: ...120 Protocol Systems Inc...