114
Protocol Systems, Inc.
Power Adapter Specifications
Low Battery Voltage and
Operation
< 7.8 V: Caution message LOW BATTERY.
< 7.6 V: Caution messages LOW BATTERY, PRINTER DISABLED and
LOW BATTERY, NIBP DISABLED.
< 7.4 V: Equipment alert VERY LOW BATTERY
< 7.3 V: Equipment alert LOW BATTERY, HEATER DISABLED (MCO
2
)
< 7.0 V: Monitor automatically turns off.
Characteristic
Specification
Protection Classifications, all Adapters
1
2
1. Per EN 60601-1 unless otherwise stated.
2. See the Protocol Systems Products and Accessories booklet for model numbers.
Type of Protection Against Electric Shock
Class I, (Protectively Earthed)
Degree of Protection Against Harmful
Ingress of Water
For ordinary, indoor locations only.
Method of Disinfection
Not suitable for autoclaving
Flammable Anesthetics
Not suitable for use with flammable anesthetics
Environmental Specifications, All Adapters
Operating Temperature
0° to 50° C
Shipping and Storage Temperature
-20° to 60° C
Operating Altitude
-2,000 to 15,000 feet (-610 to 4,572 m)
Shipping and Storage Altitude
-2,000 to 40,000 feet (-610 to 12,192 m)
Operating Relative Humidity
15% to 95%, noncondensing
Shipping, Storage Relative Humidity
15% to 95%, noncondensing
Shock
50 g
Vibration
Random Vibration, 0.02 g
2
/Hz from 10 to 300 Hz, ramping down
to 0.002 g
2
/Hz at 500 Hz. Operating 1 hour per axis, 3 hours/
test.
Physical Specifications
Length
5.0 in (12.7 cm)
Width
3.6 in (9.1 cm)
Height
3.1 in (7.9 cm)
Weight
3.1 lb (1.4 kg)
Universal Power Adapter, Part No. 503-0054-00
Rated Input
100-120 V ac, 500 mA, 50/60 Hz
Rated Fuses
T800 mA/250 V, Time-Delay, 5x20mm
Rated Output (Continuous)
16-24 V dc, 25 VA
Additional Features
Detachable power cord, pilot light
Universal Power Adapter, Part No. 503-0054-01
Rated Input
200-240 V ac, 250 mA, 50/60 Hz
Rated Fuses
T400 mA/250 V, Time-Delay, 5 x 20mm
Rated Output (Continuous)
16-24 V dc, 25 VA
Additional Features
Detachable power cord, pilot light
Characteristic
Specification
Power Adapter
connector
Positive (+)
Negative (-)
Summary of Contents for Propaq 242
Page 6: ...6 Protocol Systems Inc...
Page 18: ...18 Protocol Systems Inc...
Page 26: ...26 Protocol Systems Inc...
Page 50: ...50 Protocol Systems Inc...
Page 52: ...52 Protocol Systems Inc...
Page 64: ...64 Protocol Systems Inc...
Page 66: ...66 Protocol Systems Inc...
Page 72: ...72 Protocol Systems Inc...
Page 74: ...74 Protocol Systems Inc...
Page 78: ...78 Protocol Systems Inc...
Page 84: ...84 Protocol Systems Inc...
Page 86: ...86 Protocol Systems Inc...
Page 94: ...94 Protocol Systems Inc...
Page 120: ...120 Protocol Systems Inc...