70
Protocol Systems, Inc.
Set Printer Options and Automatic Printing
1.
Press
SETUP, MORE, PRINTER
from the Main Menu to display the Printer Setup Page:
2.
Press
NEXT
and
CHANGE
as desired to set printer options.
CONTINUOUS
Set the speed for continuous printing: 6.25, 12.5, or 25.0 mm/s.
AUTO PRINT
Automatically print a waveform snapshot at the specified interval: 15 or 30
minutes, or 1, 2, or 4 hours (or OFF).
ALARM PRINT
If ON, automatically prints patient data whenever a patient alarm occurs,
beginning with 12 seconds of patient data history stored before the alarm
occurred.
Printing continues for 20 seconds after you suspend the alarm. To immediately
stop printing, press
START/STOP
.
Note
Because the Alarm Print begins with the 12 seconds of patient data stored before the alarm occurred, the monitor
stores and prints all Alarm Print data 12 seconds after the patient data appears on the display. The time annotated on the Alarm
Print indicates the time the data was recorded.
NIBP TICKET
If ON, automatically prints an NIBP TICKET with NIBP data whenever
NIBP is measured.
APNEA TICKET
If ON, automatically prints an APNEA TICKET with apnea data after the
patient resumes breathing and/or every minute the apnea alarm continues.
OXYCRG ON ALARM
If ON, automatically prints an OxyCRG whenever an SpO
2
, HR/PR,
RR/BR, or apnea patient alarm occurs (see page 71).
If an SpO
2
or HR/PR alarm occurs, the OxyCRG prints 60 seconds later. If an
Apnea or RR/BR alarm occurs, the OxyCRG prints 75 seconds later. Highlighted
labels in the printout indicate which alarms occurred.
AUTO TREND
Automatically print trends at the selected hours. Only the parameters set to
ON (for NIBP, RESP, P1, P2, or TEMP) are included in trend printouts.
PRINTER
BR
TEMP
100.4
98.6
1.8
97
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
°
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
S
D
NEXT
T1
T2
∆
T
PREVIOUS
MENU
SETUP PAGE
AUTO TREND : OFF
CONTINUOUS
AUTO PRINT
ALARM PRINT
NIBP TICKET
APNEA TICKET
OXYCRG ON ALARM
NIBP : ON
RESP : OFF
P1
: ON
:
25.0 mm/s
: OFF
: OFF
: OFF
: ON
: OFF
P
2
: OFF
CHANGE
TREND
TEMP : OFF
12
Press to print all
selected trends.
Press to scroll to
the next selection.
Press to change the
displayed value.
Summary of Contents for Propaq 242
Page 6: ...6 Protocol Systems Inc...
Page 18: ...18 Protocol Systems Inc...
Page 26: ...26 Protocol Systems Inc...
Page 50: ...50 Protocol Systems Inc...
Page 52: ...52 Protocol Systems Inc...
Page 64: ...64 Protocol Systems Inc...
Page 66: ...66 Protocol Systems Inc...
Page 72: ...72 Protocol Systems Inc...
Page 74: ...74 Protocol Systems Inc...
Page 78: ...78 Protocol Systems Inc...
Page 84: ...84 Protocol Systems Inc...
Page 86: ...86 Protocol Systems Inc...
Page 94: ...94 Protocol Systems Inc...
Page 120: ...120 Protocol Systems Inc...