Used calibration solution
Dispose of the used calibration solution. For more
information: see calibration solution safety data
sheet.
Two test containers with a calibration solution are required for cali‐
bration. The fluoride content of the calibrating solutions should be
at least 0.5 ppm F
-
apart from each other. The sensor should be
rinsed thoroughly with fluoride-free water when changing the cali‐
brating solution.
1.
Press the key in the continuous display.
2.
Using the arrow keys select
[Two point calibration]
3.
Then press
CAL F-
Immerse sensor in buffer 1
Two point calibration
A1038
Sensor value
2.50 ppm
Sensor voltage
161.4 mV
Start with <CAL>
Fig. 53: Fluoride sensor calibration (CAL)
4.
Immerse the sensor in test container 1 with calibration solu‐
tion. When doing so gently move the sensor
5.
Then press
ð
[Calib. in progress] .
CAL F-
Two point calibration
A1040
Sensor value
2.50 ppm
Change with <OK>
continue with <CAL>
Fig. 54: Fluoride sensor calibration (CAL)
6.
Then press to change the ppm value or press to con‐
tinue with the calibration
7.
Then press
Calibrating the controller diaLog DACb
90