CAL pH
Last calibration
A1016
Zero point
Slope
Calibration process
CAL setup
continue with <CAL>
14:26:07
7,00 pH
2 point
56.64mV/pH
06/04/2013
Fig. 37: pH sensor calibration (CAL)
1.
Continue with .
2.
Rinse the sensor thoroughly with water and dry with a cloth
(pat dry, don't rub).
3.
Immerse the sensor in test container 1 which contains the
buffer solution (e.g. pH 7). Gently move the sensor.
4.
Continue with .
Do not move the sensor cable during calibra‐
tion as this can lead to signal variations.
ð
Calibration is running .
[Please wait!] flashes.
Sensor voltage
Sensor calibration in buffer 1
A1017
Buffer temperature
Stability
acceptable
good
Continue with <CAL>
Fig. 38: Display of the sensor stability achieved
5.
The range
[acceptable / good / very good] is displayed.
ð
The black part of the horizontal bar indicates the deter‐
mined range.
6.
As soon as the black bar appears, the display changes from
[Please wait!] to continue with .
The black bar does not need to be at
[very good].
Calibrating the controller diaLog DACb
77