CAL pH
Calibration process Data input
A1024
Zero point
Slope
Last calibration
CAL setup
continue with <CAL>
16:47:32
7,00 pH
59.16 mV/pH
06/05/2013
Fig. 43: pH sensor calibration (CAL)
1.
Continue with .
CAL pH
6.88 pH
Slope
at 25.0 °C
A1026
or
-58.07 mV/pH
continue with <CAL>
Asymmetry
at 25.0 °C
Zero point
at 25.0 °C
-6.4 mV
Fig. 44: Selection of the settable parameters
2.
Use the arrow keys to select the required menu entry and
press .
ð
The Enter window appears.
3.
Use the arrow keys to enter the values of your sensor and
press .
4.
Continue with .
Incorrect calibration
An error message appears if the result of the cali‐
bration lies outside the specified tolerance limits. In
this case the current calibration is not carried over.
Check the prerequisites for calibration and elimi‐
nate the error. Then repeat calibration.
5.
Carry over the result of the calibration into the controller
memory by pressing .
ð
The controller shows the continuous display again and
operates with the results of the calibration.
Calibrating the controller diaLog DACb
83