Continuous display ➨
CAL ORP
CAL setup
A1032
Offset
21.05.2013 14:59:56
continue with <CAL>
Calibration process
Last calibration
0.0 mV
Data input offset
Fig. 50: Data input, ORP sensor calibration (CAL)
1.
Continue with
CAL ORP
Offset
A1033
Accept with <CAL>
0.1 mV
Fig. 51: Adjusting the [Offset]
2.
Press and use the four arrow keys to adjust the mV value
of the buffer you are using
3.
Press
4.
Transfer the result of the calibration into the controller
memory by pressing
ð
The controller operates with the calibration results.
15.1.3 Calibrating the Fluoride Sensor
15.1.3.1
Selection of the calibration process for fluoride
To calibrate the controller there are two available calibration pro‐
cesses:
n
1 point
n
2 point
Calibrating the controller diaLog DACb
88