16.1.2 Calibration of the temperature coefficient
Conductivity sensors with temperature element
You can only calibrate the temperature coefficient
with conductivity sensors with a temperature ele‐
ment, because it is impossible to calculate the tem‐
perature coefficient without measuring the tem‐
perature.
Temperature change
It is recommended that the temperature is changed
by no more than 0.5 °C per minute, or with a tem‐
perature change of e.g. 10 °C you will need to wait
for a minimum of 20 minutes before calibration.
A1866
CAL
CELLCONST
CAL TEMPCOEFF
CAL_T1= 22.0 °C
CHANGE TEMP
TCOEFF
CAL TEMPCOEFF
T2 > 32.0 °C or
CAL_T2= 43.0 °C
CAL TEMPCOEFF
TCOEFF=0.00%/°C
CAL TEMPCOEFF
CAL=OLD
CAL=CONTINUE
T2 < 12.0 °C
CAL=CONTINUE
WAIT
CAL TEMPCOEFF
T2 > 32.0 °C or
T2 < 12.0 °C
I.
II.
STABLE ?
CAL_T2= 43.0 °C
CAL TEMPCOEFF
CAL=CONTINUE
ACCEPT ?
III.
Fig. 80: Calibration of the temperature coefficient
I. If the temperature change is greater than 2°C,
the message changes to
[WAIT]
II. If the temperature is within the specified range,
the message changes to
[STABLE ?]
III. When a stable final temperature has been
reached, the message changes to
[ACCEPT?].
Calibration can now be terminated manually.
Calibrating the compact controller for inductive conductivity
115