7.
Wait for the stability of the measured values.
8.
The display appears as follows if calibration is successful:
CAL O2
A2854
Before calibration
Slope
Calibration successful
Cancel with <ESC>
After calibration
A
Zero point
End with <CAL>
Fig. 67: Successful calibration
9.
n
Use CAL to confirm.
n
Press ESC to cancel.
10. The display appears as follows if calibration is unsuccessful:
CAL O2
A2855
Before calibration
Slope
Calibration not possible
Cancel with <ESC>
After calibration
A
Zero point
End with <CAL>
---
i
Fig. 68: Unsuccessful calibration
11.
n
Use CAL to end.
n
Press ESC to cancel.
12. Check the sensor and installation once again and repeat the
calibration procedure.