CAL O2
A2859
Determine reference value
Change with <OK>
Continue with <CAL>
07.51 ppm
Fig. 71: Reference value
6.
Use OK and the arrow keys to enter the reference value
ð
The display appears as follows:
CAL O2
A2858
Reference value
Range:
0.00 ... 24.00 ppm
07.51 ppm
Fig. 72: Reference value
7.
Continue with CAL.
8.
The display appears as follows if calibration is successful:
CAL O2
A2854
Before calibration
Slope
Calibration successful
Cancel with <ESC>
After calibration
A
Zero point
End with <CAL>
Fig. 73: Successful calibration
9.
n
Use CAL to confirm
n
Press ESC to cancel
10. The display appears as follows if calibration is unsuccessful:
Calibrating the controller diaLog DACb
104