6 PRODIGIT
CP Mode:
At Constant Power mode, the 3350F Series Electronic Load will attempt to sink load power
(load voltage x load current) in accordance with the programmed power. (See Fig 1-10).
Fig 1-10 Constant Power mode
Dynamic wave form definition:
There are six parameters to generate dynamic wave form or pulse wave form, the 3350F
Series Electronic Load will sink current from power source proportional to the dynamic wave
form, the dynamic wave form definition is shown in Fig 1-11. The period of dynamic wave form
is Thigh + Tlow, dynamic frequency = 1 / (Thigh + Tlow), the Duty cycle = Thigh / (Thigh + Tlow)
I
V
LOAD
CURRENT
RISE
SLEW RATE
HIGH LOAD
LOAD
LOW LOAD
LEVEL
T
HIGH
T
LOW
FALL SLEW RATE
Fig 1-11 Dynamic Wave form