42 PRODIGIT
Setting the short test time
,
The LCD display show
「
SHORT
」
on left 5 digits
LCD display
,
shows
「
TIME
」
on middle 5 digits LCD display
,
right 5 digit LCD
display
「
CONTI
」
, the unit is "ms".
TIME: setting the short test time, The LCD display show
「
SHORT
」
on left 5
digits LCD display, shows
「
TIME
」
on middle 5 digits LCD display the unit is
“ms” ,and shows
「
CONTI
」
on right 5 digits LCD display, the Setting range is
“CONTI” means continue, 100mS to 10000mS step 100ms by Clockwise rotate the
Setting knob.
The short test will be no time limitation when setting to CONTI until press
“START/STOP” key to stop the short test.
V-Hi : Short test voltage check upper limitation setting, The LCD display shows
「
SHORT
」
on left 5 digit LCD display, Middle 5 digit LCD display
「
V-Hi
」
,right 5
Digit LCD display setting value, the unit is "V", The V-Hi setting range from 0.000V to
60.000V step 0.001V by rotating the setting knob.