4.6.3. STAGE Set and read the status of Load
[STATe
:
] LOAD {SP}{ON
OFF}
Syntax
:
[STATe
:
] LOAD{SP}{ON
OFF}{
;
NL}
[STATe
:
] LOAD
?
{
;
NL}
Purpose
:
Set and read the status of Sink Current or not
Description
:
This command is used for setting the status of Sink Current . When
setting it to ON, the Load is going to sink current from appliance. When
setting it to OFF, the Load would not act.
[STATe
:
] MODE {SP}{CC
CR
CV
CP}
Syntax
:
[STATe
:
] MODE {SP}{CC
CR
CV
CP}{
;
NL}
[STATe
:
] MODE
?
{
;
NL}
Purpose
:
Set and read the mode of LOAD
Description
:
Load is acting under these four modes as the following TABLE 4-9. When
reading the Loading Operation mode, the return value 0│1│2│3 are meant
to be CC│CR│CV│CP
CC
(0)
CR
(1)
CV
(2)
CP
(3)
3350F
V
V
V
V
Table 4-9 module for each series
[STATe
:
] SHORt {SP}{ON
OFF}
Syntax
:
[STATe
:
] SHORt {SP}{ON
OFF}{
;
NL}
[STATe
:
] SHORt
?
{
;
NL}
Purpose
:
Set and read the short-circuit test of Load
Description
:
This command is for setting the Load to make a short-circuit test. While
setting for the ON, the V+, V- pin of Load like short-circuit status.
[STATe
:
] PRESet {SP}{ON
OFF}
Syntax
:
[STATe
:
] PRESet {SP}{ON
OFF}{
;
NL}
[STATe
:
] PRESet
?
{
;
NL}
Purpose
:
Set the upper or lower digit multi-function meter to display the programming
load level.
Description
:
This command is for select the left 5 digit LCD display to show current
setting or DWM.
Pres ON
:
To select the LCD display to shows current setting
Pres OFF
:
To select the LCD Display is
〝
DWM
〞