28 PRODIGIT
3.2.7.4. CONFIG. Each press of the CONFIG button changes the middle LCD
Text.
The sequence and the corresponding setting value shown on the bottom
Display are as follows:
SENSE can be set to
「
AUTO
」
or
「
ON
」
LDon (load ON voltage) displays the set value in volts “V”
LDoff (load OFF voltage) displays the set value in volts “V”
POLAR (load polarity) can be set to
「+
LOAD
」
or
「-
LOAD
」
3.2.7.5. SHORT test. This allows the parameters of the short test to be set up.
Each press of the SHORT button moves the setting function. The
Sequence of the short test along with the setting value is as follows:
Short Press Start (pressing the red START/STOP button starts the test)
TIME shows the duration of the SHORT test. “CONTI”, on the bottom
display indicates continuous. Time can be adjusted in “ms”.
V-Hi (voltage high threshold) displays the set value in volts “V”
V-Lo (voltage low threshold) displays the set value in volts “V”
When the test is started the lower display will show RUN. When the test
Has finished the lower display will show END.
3.2.7.6. OPP test. This allows the parameters of the over power protection test to
Be Set up. Each press of the OPP button moves the setting function. The
Sequence of the OPP test along with the setting value is as follows:
OPP Press Start (pressing the red START/STOP button starts the test)
PSTAR (power start point) right display provides setting in watts “W”
PSTEP (power steps) right display provides setting in watts “W”
PSTOP (power stop point) right display provides setting in watts “W”
VTH (voltage threshold) right display provides setting in volts “V”
When the test is started the right display will show the power value
Being taken by the load. If the Device under Test is able to supply the
Load according to the values set then the middle display will show PASS
And the right display will show the maximum power taken during the
OPP test. If, during the test, OTP is displayed the over temperature
Protection has been engaged. Similarly if OPP is shown on the display
The over power protection has been activated.
3.2.7.7. OCP test. This allows the parameters of the over current protection test
To be set up. Each press of the OCP button moves the setting function.
The sequence of the OCP test along with the setting value is as follows:
OCP Press Start (pressing the red START/STOP button starts the test)
ISTAR (current start point) right display provides setting in amps “A”
ISTEP (current steps) right display provides setting in amps “A”
ISTOP (current stop point) right display provides setting in amps “A”
VTH (voltage threshold) right display provides setting in volts “V”