82 PRODIGIT
Measure command: available for all module
COMMAND
RETURN
MEASure
:
CURRent{?}{
;
NL}
###.####
MEASure
:
VOLTage{?}{
;
NL}
###.####
MEASure
:
POWer{?}{
;
NL}
###.####
Table 4-6B MEASURE COMMAND SUMMARY
REMARK
:
1. Current engineering unit: A
2. Voltage engineering unit: V
3. Period engineering unit: mS
4. Slew-rate engineering unit: (m)A/uS
5. Power engineering unit: W
Auto sequence: avilable for all module
AUTO SEQUENCE COMMAND
NOTE
RETURN
FILE {SP} {n}{
;
NL}
n=1~9
1~9
STEP {SP} {n} {
;
NL}
n=1~16
1~16
TOTSTEP {SP} {n}{
;
NL}
Total step n=1~16
1~16
SB {SP} {m,n} {
;
NL}
m=1~10 n=1~15
m:STATE , n:BANK
T1 {SP} {NR2} {
;
NL}
0.1~9.9(s)
0.1~9.9(sec)
T2 {SP} {NR2} {
;
NL}
0.0~9.9(s)
0.0~9.9(sec)
SAVE {
;
NL}
Save “File n” data
REPEAT {SP} {n} {
;
NL}
n=0~9999
0~9999
RUN {SP} {F} {n} {
;
NL}
n=1~9
AUTO REPLY
“PASS” or “FAIL:XX”
(XX=NG STEP)
Table 4-8B Auto sequence command list