3302F Series Mainframe Operation Manual 51
Purpose:
Set and read the BW level to LO or HI.
Description
: The BW command allows the user to changes the bandwidth of the
33401F/G / 3340F/G series load when operating in LED mode. The BW
Function can be Used when in CC Mode, CV Mode and CR Mode Range
I. The initial Value of the bandwidth is HI.
Please note that when operating in CR Mode Range II there is only one bandwidth,
AVG
Command Syntax
[PRESet
:
] AVG{SP}{n}
[PRESet
:
] AVG{?}
Purpose:
set and read VI Measuring Average
Description
: A number of voltage and current measurements can be taken and their
Values averaged. The average function can be set in the range of 1 to 64.
The default value is 1.
LEDNO
Command Syntax
[PRESet
:
] LEDNO {SP}{n}
[PRESet
:
] LEDNO {?}
Purpose:
Set and read quantity of LEDs to be simulated
Description
: This command allows the user to enter the total number of LEDs the
33401F/G /3340F/G series is to simulate.
•
For models 3340F/G and 3342F/G the LEDNO setting range is 1 to 90
•
For the 3341F/G model the LEDNO setting range is 1 to 30
•
For the 33401F/G model the LEDNO setting range is 1 to 90
VO
Command Syntax
[PRESet
:
]VO
{
SP
}
{
NR2
}
{
;
NL
}
[PRESet
:
]VO {?}{;|NL}
Purpose:
Set and read
Vo Voltage in LED mode.
Description
: This command is used to set and read back the output voltage VO for
The 3340F/G series of LED simulators. The engineering unit is V
VD
Command Syntax
[PRESet
:
]VD
{
SP
}
{
NR2
}
{
;
NL
}
[PRESet
:
] VD {?}{;|NL}
Purpose:
Set and read
Vd Voltage in LED mode.
Description
: This command is used to set and read back the forward bias voltage (Vd)
Of the 3340F/G units when operating in LED mode. The engineering unit
is V.
RD
Command Syntax
[PRESet
:
] RD
{
SP
}
{
NR2
}
{
;
NL
}
[PRESet
:
] RD {?}{;|NL}
Purpose:
Set and read RD Impedance
in LED mode.