46 PRODIGIT
PERI or PERD
Command Syntax
:
[ PRESet
:
] PERI
PERD
:
HIGH
LOW{SP}{ NR2}{
;
NL}
[ PRESet
:
] PERI
PERD
:
HIGH
LOW
?
{
;
NL}
Purpose:
Set and read the combined TLOW and THIGH of a DYNAMIC
Waveform
Description:
1. The time period combines TLOW (time low) and THIGH (time high) sections of
a DYNAMIC waveform.
2. The value of TLOW and THIGH has to be included the number of the decimal
point; otherwise the command will not be available.
3. The least significant number is the 5th after the decimal point.
4. Should a value be entered that is higher than what is possible then the 3302F
will automatically set its maximum value according the load module fitted
5. The engineering unit is ms.
LDONv
Command Syntax
:
[ PRESet
:
] LDONv {SP}{NR2}{
;
NL}
[ PRESet
:
] LDONv?{
;
NL}
Purpose:
Set and Read the voltage level which the LOAD will switch ON
Description:
This command is for setting the voltage value at which the LOAD
Will automatically switch ON. The engineering unit is V.
LDOFfv
Command Syntax
:
[ PRESet
:
] LDOFfv{SP}{ NR2}{
;
NL}
[ PRESet
:
] LDOFfv? {
;
NL}
Purpose:
Set and Read the voltage level which the LOAD will switch OFF
Description:
This command is for setting the voltage value at which the LOAD will
automatically switch OFF. The engineering unit is V.
CURR
:
HIGH
LOW
Command Syntax
:
[ PRESet
:
] CC
CURR
:
HIGH
LOW{SP}{ NR2}{
;
NL}
[ PRESet
:
] CC
CURR
:
HIGH
LOW
?
{
;
NL}
Purpose:
Set and read the HIGH
LOW current levels.
Description:
This command is used for setting the HIGH and LOW levels of
Load current. These 2 current levels need to be used should a dynamic load
Waveform is desired. It also allow the user to switch between 2 preset current
Levels.
1. The least significant number is the 5
th
digit after the decimal point.
2. The LOW level current value cannot be higher than the HIGH level.
3. Should a value be entered that is higher than what is possible then the
3302F will automatically set its maximum value according the load
module fitted.
4. The engineering unit is A.