HP-MSM-HB
35,0(6
NOTICE
Damaging/destroying the device
Contamination and fingerprints on the protective window / mirror can lead to damage or
shattering or splintering of the protective window during measuring operation.
X
Only replace the protective window in a dust-free environment.
X
Do not touch the protective window.
X
When exchanging the protective window wear powder-free latex gloves.
15.1
Demounting the Measuring Objective
1. Turn off the voltage supply.
2. Turn off the cooling water and the compressed air supply.
3. Mount the transport lock at the beam entrance aperture.
4. Disconnect the cooling water pipes at the quick connect couplings.
5. Remove compressed air hose from the connection for the housing flushing.
6. Remove the fastening screws(see chapter 3.3.4 on page 13).
7. Remove measuring objective
8. Drain cooling water and blow the cooling water pipes out with compressed air.
9. Dry and clean the measuring objective.
Beam Entrance
Measuring Output
Laser Output
Fig. 15.1:
Openings of the measuring objective