54
Revision 00 EN - 01/2021
HP-MSM-HB
35,0(6
13.6
Perform measurement settings in the LaserDiagnosticsSoftware LDS
The following explanations of the configuration options should help you to make the right settings for the
respective task.
The following chapters highlight important configuration options in color:
Color
Meaning
Red
This setting must always be set as shown.
Yellow
This setting is dependent from the desired operating mode (CW, pulse, single pulse, measurement
series, etc.).
Green
This setting must be carried out before each measurement. The settings depend on the specific mea-
surement task, such as the wavelength, the laser power or the geometry of the laser beam.
Tab. 13.4: Color meaning of the setting options
13.6.1
Sensor parameters (menu
Measurement > Sensor parameter
)
Adaptable squares
Z
Y
Fig. 13.2:
Dialogue window
Sensor parameters
Mechanical limits
By pulling the turquoise square with the mouse pointer you can restrict the movement range of the y- and z-
axis. Therewith you can prevent damages in case other components reach into the movement range of your
device. The maximum value corresponds to the value Y3 and Z3.
Resolution
Here you can enter the number of pixels in the measuring window, ranging from 32 x 32 to 256 x 256 pixels.
Generally, 64 x 64 pixels are sufficient. Please keep in mind that the more pixels there are, the longer the
measurement will take.