![Primes HP-MSM-HB Original Instructions Manual Download Page 122](http://html1.mh-extra.com/html/primes/hp-msm-hb/hp-msm-hb_original-instructions-manual_1601727122.webp)
122
Revision 00 EN - 01/2021
HP-MSM-HB
35,0(6
For TopHat and Gaussian beam shaped laser beams, the fill factor should range between 0.5 and 0.7. If the
beam has diffraction rings, however, and if these are located completely within the measuring window, the
optimal value for the fill factor can be between 0.5 and 0.6.
By default, the value should be set to: “Max 0.7 Min 0.5 Target 0.6”. For extremely deformed beams, the
value may be changed to “Max 0.6 Min 0.4 Target 0.5”.
Font size
The font size for the most important display window can be changed here.
It is set to 10 points at the factory.
Open windows
When the window opening function is activated, some basic windows are opened when the LDS is started. If
you don’t want this to happen, the function can be deactivated.