24.3.2
Sensor parameters (menu
Measurement > Sensor parameter
)
Adaptable squares
Z
Y
Fig. 24.5:
Dialogue window
Sensor parameters
Mechanical limits
By pulling the turquoise square with the mouse pointer you can restrict the movement range of the y- and z-
axis. Therewith you can prevent damages in case other components reach into the movement range of your
device. The maximum value corresponds to the value Y3 and Z3.
Device
By means of this option, you can select the device which is supposed to be operated. Depending on the
number of devices connected, additional device numbers are assigned.
RPM
Not relevant for HP-MSM-HB.
Resolution
Here you can enter the number of pixels in the measuring window, ranging from 32 x 32 to 256 x 256 pixels.
Generally, 64 x 64 pixels are sufficient. Please keep in mind that the more pixels there are, the longer the
measurement will take.
Detector
Not relevant for HP-MSM-HB.
Manual z-axis
With this function you can deactivate the z-axes of the measuring system. This is useful if you want to use
external movement axes. In this case you can manually assign a z-value to every measurement plane in the
dialogue window
Single measurement
.
24.3.3
Beam find settings (menu
Measurement > BeamFind Settings: Beamfind