15.1
Changing the Detector
You just have to open the inspection opening in the bottom of the housing which is fastened by means of
four Torx screws (T8).
T8
Fig. 15.1: Inspection opening of the BeamMonitor BM+
NOTICE
Danger of damage for the detector sensor
The detector sensor must not be damaged and has to be protected from contaminations
of any kind.
Do not touch the detector sensor with your fingers and do not put it down on the sensor
surface.
Sensor surface
i
Please only use isolating plastic screws in order to fasten the detector in order to prevent noise
signals. Do not forget the foam rubber plate during the installation. Otherwise the rotational disc
could be mechanically blocked by the screws. The foam rubber plate also ensures a mechanical
decoupling and electrical isolation of the detector.