PRIMES
BeamMonitor BM+ with LDS 2.98
29
Revision 01/2018 EN
192 168 116
84
Fig. 10.6: Establishing a connection in the menu
Free Communication
10.4.2
Changing the IP address
You can change the preset IP address in the menu
Communication>>Free communication
by means of
the following commands:
IP-address
(Sample address)
192.
168.
116.
84
Commands
se0328
xyz
se0329
xyz
se0330
xyz
se0331
xyz
In this case
xyz are place holders of the four IP-address bytes (values 1 - 255) which always have to be
entered with three digits!
For example, the number 84 has to be entered like this: 084.
For reasons of clarity the symbol
marks a space.
Example:
You will change the IP address from 192.168.116.85 to 192.168.116.
86.
1. Please start the PRIMES LaserDiagnosticsSoftware.
2. Open the menu
Communication>>Free Communication
.
3. Choose the mode “TCP” (the option “second IP” must not be activated!).
4. Enter the IP in the field “TCP”.
5. Click on the “connect” button (“connected” appears in the bus monitor).
6. Activate the check box
Write bus protocol
(the protocol can be helpful in case of problems).
7. Enter the following in the field
Command
(please make sure that the blank character
is entered cor-
rectly):
se0331
086
8. Click to
Send
and wait for the confirmation in the bus monitor (in Fig. 10.7 „-> Adr:0331 Wert: 086“)
9. Please turn off the device and turn it on again. After this restart the IP-address is updated.