7.4
Mounting
DANGER
Serious eye or skin injury due to laser radiation
If the appropriate position of the measuring device is changed, this could cause increased
scattered radiation during the measurement.
When mounting the device, please ensure that it cannot be moved, neither due to an unin-
tended push or a pull on the cables and hoses.
NOTICE
Damage/Destruction of the device
Due to screws which are too long, internal components could be destroyed.
Please consider, that the fastening screws must only extend up to 10 mm into the housing.
There are four tapped holes M6 in the mounting surface of the housing intended for the customer’s mount-
ing. Please use at least four screws to fasten the device. The total length of the screws depends on the
dimensions of the customer’s mounting. The dimensioned order of the fixing holes can be found in chapter
19 on page 73.