8.2
Operation of two measuring devices
Power On/Off
Power In
BM+
BeamMonitor
Ethernet
PRIMES-Bus
RS 485
Ethernet
PC
BeamMonitor BM+
PRIMES Power Supply
Error
Power
Interlock
PRIMES-Bus
RS 485
RS 485
RS 485
Analog
Out
USB
Clean
Compressed Air
Water Out
Water In
PM100
PowerMonitor
PowerMonitor 100
PRIMES Bus
RS485
Fig. 8.3:
Connection of the BeamMonitor BM+ and the PowerMonito
i
For connection of several devices please use only one power supply (typically PRIMES power
supply).
NOTICE
There is a danger of damage
When disconnecting the bus connections during the operation (when the system is con-
nected with the supply voltage), voltage peaks can develop which could destroy communi-
cation modules of the measuring devices.
Please turn off the power supply before disconnecting the bus cables.