PRIMES
BeamMonitor BM+ with LDS 2.98
63
Revision 01/2018 EN
5. Please open the dialogue window
Measurement>>Single...
and select
C Window size in x-direction: recommended setting X=0.8 mm
Window size in y-direction: recommended setting Y=0.8 mm
D False colors
C
D
E
F
i
E In the range “Power” the laser power of the beam which is to be measured has to be typed in.
F Please turn on the laser and click on the “Start” button.
i
With regard to the BeamMonitor BM+ the button
Beam Find
automatically deals with the positioning
and the selection of the measuring window. With regard to this, the z-position remains unchanged
and the search is limited to the window range. In case the search was successful, the measuring
window found is displayed. It is only a search function. The measuring window size is not optimized.
Signal saturation (for detectors of NIR- and CO
2
-Lasers):
In case the signal is too big, the amplification can be reduced.
Detector type
Adjustable sensitivity
DBY-PS+
Yes (automatically)
DBIG-PS+
Yes (automatically)
DBCM+
No