11.2.11 Graphical Review
The display window
Graphical review
offers many possibilities to display the measurement values of
the single measurement planes.
Above the x-axis the power, time and planes or the z-position can be applied. For the y-axis the radius data,
the x or, respectively, y-position, the angle and the ellipticity are available. In total this window can present
16 different graphs.
Fig. 11.25: Graphical review - example radius versus plane
11.2.12 Color Tables
Different color charts are available. It is possible to switch back and forth between the color charts. Thus the
assignment of A/D converter values and different color scales can be varied. This is important for the false
color presentation.
Three settings are possible:
•
Linear color table (basic setting)
•
Color table analogue to the root function
•
Color table analogue to the fourth root function
These functions can especially be helpful as far as the analysis of slight variations near the zero level are
concerned; e.g. the analysis of diffraction phenomena.