12.3
Possible Types of Measurement
12.3.1
Single Measurement
Only one measurement in one plane is carried out. The single measurement can be adjusted automatically
or manually. The position and the size of the measuring window can be adjusted relatively to the maximum
measurement range. The amplification is adjustable separately. A false color presentation is possible.
i
When measuring with the DBY-PS+ Detector you have to carry out a manual single measurement
before the caustic measurement can be started (please see chapter 12.5 on page 65).
12.3.2
Caustic measurement
Several measurements in different planes of the z-axis are carried out. The parameters can be adjusted
automatically or manually in the menu item
Measurement, Measurement Settings
. The measurement
enables the direct determination of the beam propagation ratio M² (beam propagation factor K).
12.4
Brief Instruction for a first Single Measurement
i
Please turn on the supply voltage of the device and wait for about 20 seconds. Then the software
can be started.
When turning the supply voltage on and off a reset cycle is started within the device. During this
time no measurements are possible!
1. Please turn on the supply voltage. Wait for approximately 20 seconds until the boot process is finished.
2. Please start the LaserDiagnosticsSoftware. The connected devices are recognized within 20 seconds
and in the upper right hand corner the device symbols are displayed.
Device symbols
In case the device is
not recognized:
Please open the dialogue window
Communication>>Free Communication
and make the following
settings:
•
Select the mode TCP (the option „Second IP“ must not be activated).
•
Type in the IP address of the device in the field „TCP“.
•
Click the button „Connect“ („Connected“ appears in the bus monitor).
•
Click the button
“
Save
”
(this configuration will be stored and needs not to be typed in after restart of
the software).