Power supply connection
Harting M12-P-PCB-THR-2PC-5P-LCOD-M-STR
Pin
Function
1
+24 V
2
Not assigned
3
GND
4
Not assigned
5
FE
Tab. 8.1:
Connection socket for the power supply
PRIMES Bus
Pole arrangement D-Sub-socket, 9 pole (view of plug-in side)
1
5
6
9
Pin
Function
1
GND
2
RS 485 (+)
3
+24 V
4
Trigger RS 485 (+)
5
Not assigned
6
GND
7
RS 485 (–)
8
+24 V
9
Trigger RS 485 (–)
Tab. 8.2:
D-Submin-socket, PRIMES-Bus
In case you would like to use self-configured cables, please keep the following aspects in mind:
•
The length of the cable reaching from the power supply to the measuring device must not exceed 1.8 m.
Otherwise the voltage drop of the cable would be too high.
•
Please use shielded cable only and observe, that the shielding is efficient continuously.
•
The cable length between the BeamMonitor BM+ and the second device (via RS485) must neither
exceed 2 m.