When the touch display is locked, the Lock button is blue. Press the button
for 1 second to unlock the touch display. A progress bar is displayed while
you press the button.
Alternatively, the dental unit can be configured so that you can lock/unlock
the touch display by pressing the Flexy button. Contact your Planmeca
dealer.
NOTE
When the dental unit is equipped with hanging-tube instruments, dry the
instrument holders properly after cleaning. Wet surfaces might disturb
instrument recognition.
NOTE
The instrument console and control panel must be completely dry before
covering them with a protective cover.
28.3 Instruments
Clean and service the instruments according to the information supplied with
the instrument.
NOTE
After cleaning the instrument, let the oil run from it for at least 10 minutes
before replacing it on the console.
Do not use lubricating oil on the Morita TORX micromotor (with the exception
of its O-rings).
Throw away the intraoral camera’s disposable hygiene sleeve after use.
28.4 Cuspidor
28.4.1 Bowl
NOTE
Do not pour anything into the bowl unless the unit is switched on and has air
and water (that is, the compressor is switched on and the air- and waterlines
are opened).
NOTE
Do not pour anything else than water and a suction line cleaning agent into
the bowl. The water must be poured slowly into the bowl. The flow may not
exceed 5 l/min.
NOTE
The bowl can be removed and washed in the dishwasher. The maximum
washing temperature is 65°C. At higher temperatures the bowl may break.
Avoid rapid temperature changes in the dishwasher and when the bowl is in
use. When positioning the bowl into the dishwasher, make sure that the bowl
does not press other objects in the machine.
Clean the bowl after every patient by pouring a few drops of the cleaning
solution into the bowl and using a soft brush. Rinse the bowl by pressing
Bowl rinse. The outside of the bowl can be wiped clean with a damp cloth.
28 Cleaning and disinfection
User's manual
Planmeca Compact i 231
Summary of Contents for Compact i Touch v2
Page 1: ...PlanmecaCompact iTouchv2 user s manual EN 10033022 ...
Page 10: ...Table of contents Planmeca Compact i User s manual ...
Page 27: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 17 ...
Page 306: ...34 Technical information 296 Planmeca Compact i User s manual ...
Page 308: ...34 Technical information 298 Planmeca Compact i User s manual ...
Page 309: ...34 3 6 Cart 34 Technical information User s manual Planmeca Compact i 299 ...
Page 313: ......