
Alteration of RPM speed rate in different ranges
Range
In steps of:
20 000 - 38 000
2 000 (e.g. 22 000, 24 000 etc....)
7. Press OK.
The pop-up window closes and the new value is displayed on the control
panel.
8. Press Program.
9. Press Torque to adjust the torque limit rate.
A pop-up window opens.
10. Adjust the torque limit rate.
The adjustment range is 10% - 95%.
11. Press OK.
The pop-up window closes and the new value is displayed on the control
panel.
12. Press Program.
13. Press Auto forward to adjust the counter-clockwise rotation time of the
micromotor in the Auto forward mode.
A pop-up window opens.
14. Adjust the Auto forward time.
The maximum time is 6 seconds and the minimum 0 seconds, and the
time can be changed in steps of 0.2 seconds.
15. Press OK.
The pop-up window closes and the new value is displayed on the control
panel.
25 Programming
182 Planmeca Compact i
User's manual
Summary of Contents for Compact i Touch v2
Page 1: ...PlanmecaCompact iTouchv2 user s manual EN 10033022 ...
Page 10: ...Table of contents Planmeca Compact i User s manual ...
Page 27: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 17 ...
Page 306: ...34 Technical information 296 Planmeca Compact i User s manual ...
Page 308: ...34 Technical information 298 Planmeca Compact i User s manual ...
Page 309: ...34 3 6 Cart 34 Technical information User s manual Planmeca Compact i 299 ...
Page 313: ......