MA 2102-A005 GB
The following table gives an overview:
Additional
input to the
instruction
code
Function
Example Description
Answer
I
I
nformation about the
instruction
FI Information
about
the instruction F
f‚R/– MCD
status’
U
U
pper limit of the
number range
SU Stop
current:
highest value
S63
L
L
ower limit of the
number range
RL Run
current:
lowest value
r1
S
Scale of current or
temperature values:
1=1:1
10=1:10
SS
Read scale value
of the stop current
s10
E
Unit of current- and
time values:
A, mA, ms, s
SE
Read unit of the
stop current
eA
?
Read the preset value
R?
Display run
current: 2.5 A
r25
<value>
Set the value
S5
Stop current
0.5 A
s5
Important:
•
If the instruction input values are faulty, the actual preset value will be the answer.
•
If an instruction code is entered which is not implemented, the instruction code is
answered by a minuscule with –character.
Example: Enter K
Æ
Answer: k–