Chapter 1. Installation
1.4.2
Hazardous Area Wiring Connections
Before installing and using the HygroPro
II
in a hazardous (classified) area, be sure to read and understand all
applicable reference materials. This includes:
• all EU or North American Standards and Directives (see Table 3 and Table 4 on page 11)
• all local safety procedures and practices
• the FM Schematic drawing (see Figure 10 below)
• this user’s manual
Note:
It is the installer’s responsibility to follow all applicable technical regulations, standards and procedures.
Figure 10: HygroPro
II
Installation Control Drawing for Division 1 and Division 2 locations (refer to 752-607)
For wiring the HygroPro
II
in hazardous (classified areas) connect the flying lead end of the IS cable assembly supplied
with the transmitter as shown in the installation control diagrams in Figure 10 and Figure 11, according to one of the
following conditions:
WARNING!
The procedures in this section must be performed only by trained technicians who have
the necessary skills and qualifications.
Summary of Contents for HydroProII
Page 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Preface x HygroProII User s Manual no content intended for this page...
Page 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Page 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Page 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Page 103: ......