Appendix A. HygroPro
II
HART
®
Field Device Specification
50
HygroPro
II
User's Manual
A.9.3 Command 132 (0x84): Set Loop Current Percentage
This command is to set the output percentage of loop 1 current.
This command sets the output to a user determined percent of the 4-20 mA output.
Examples: 0 pct = 4 mA, 50% = 12 mA, and 100% = 20 mA.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0
Unsigned-8
Enumeration. 0 = Fixed percent test, 1 = Live mA
Sets live analog output or fixed mA value
1~4
Float
Loop Current Percentage, units of percent.
Byte
Format
Description
0
Unsigned-8
Enumeration. 0 = Fixed percent test, 1 = Live mA
Sets live analog output or fixed mA value
1~4
Float
Loop Current Percentage, units of percent.
Code
Class
Description
0
Success
No Command-Specific Errors
1-2
Undefined
3
Error
Passed Parameter Too Large
4
Error
Passed Parameter Too Small
5
Error
Too Few Data Bytes Received
6
Error
Device-Specific Command Error
7
Error
In Write Protect Mode
8
Undefined
9
Error
Incorrect Loop Current Mode or Value
10
Undefined
11
Error
Loop Current Not Active
12 ~ 15
Undefined
16
Error
Access Restricted
17-31
Undefined
32
Error
Busy
33 ~ 127
Undefined
Summary of Contents for HydroProII
Page 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Preface x HygroProII User s Manual no content intended for this page...
Page 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Page 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Page 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Page 103: ......