HygroPro
II
User's Manual
47
Appendix A. HygroPro
II
HART
®
Field Device Specification
A.8 Common-Practice Commands
A.8.1 Supported Commands
A.8.2 Burst Mode
This Field Device does not support Burst Mode.
A.8.3 Catch Device Variable
This Field Device does not support Catch Device Variable.
A.9 Device-Specific Commands
The following device-specific commands are implemented:
•
Command 130 (0x82): Set Loop Current Zero
•
Command 131 (0x83): Set Loop Current Gain
•
Command 132 (0x84): Set Loop Current Percentage
•
Command 133 (0x85): Set Primary Variable Range Values
•
Command 134 (0x86): Set Loop Current Error Handling
•
Command 140 (0x8C): Read Primary Variable Range Values
•
Command 141 (0x8D): Read Loop Current Error Handling
•
Command 144 (0x90): Set Loop Measurement Type
•
Command 145 (0x91): Enter / Exit Fixed Loop Current
•
Command 146 (0x92): Set Pressure Constant
•
Command 147 (0x93): Read Pressure Constant
•
Command 148 (0x94): Set Temperature Constant
•
Command 149 (0x95): Read Temperature Constant
•
Command 150 (0x96): Set Dew Point Offset
•
Command 151 (0x97): Read Dew Point Offset
•
Command 152 (0x98): Set Hygro Curve Number of Points
•
Command 153 (0x99): Read Hygro Curve Number of Points
•
Command 154 (0x9A): Set Hygro Curve
•
Command 155 (0x9B): Read Hygro Curve
•
Command 156 (0x9C): Set Pressure Curve
•
Command 157 (0x9D): Read Pressure Curve
•
Command 158 (0x9E): Set Saturation TableNumber of Points
Command
Function
Description
33
Read Device Variables
Allows a Master to request the value of up to four Device Variables.
50
Read Dynamic Variable
Assignments
Reads the Device Variables assigned to the Primary, Secondary,
Tertiary, and Quaternary Variables.
51
Write Dynamic Variable
Assignments
Allows the user to assign Device Variables to the Primary,
Secondary, Tertiary, and Quaternary Variables
54
Read Device Variable
Information
Get device variable information
59
Write Number of Response
Preambles
Sets the number of asynchronous preamble bytes to be sent by
the meter before the start of a response message.
Summary of Contents for HydroProII
Page 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Preface x HygroProII User s Manual no content intended for this page...
Page 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Page 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Page 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Page 103: ......