HygroPro
II
User's Manual
33
Chapter 3. Service and Maintenance
Table 7: List of error codes displayed HygroPro
II
screen
3.3 Cleaning the Moisture Probe
To clean your HygroPro
II
moisture probe (i.e., Hygro
II
RTE), carefully follow the instructions in this section.
3.3.1
Preparing to Clean the Probe
To clean the moisture probe, the following items are required:
•
Two glass (
NOT
metal) containers with approximately 300 ml of reagent-grade Hexane or Toluene in each
container.
•
One glass (
NOT
metal) container with approximately 300 ml of distilled (NOT deionized) water.
IMPORTANT:
Make sure the containers are deep enough to completely submerge the moisture probe after it is
disconnected from the display head. Do not place the transmitter module into any of these solvents.
Insert only the moisture sensor mounted on the probe into the solvents.
•
Rubber or latex gloves
•
An oven set at 50°C ± 2°C (122°F ± 3.6°F)
•
1-1/8” wrench
3.3.2 Replacing the RTE
To maximize HygroPro
II
performance, Panametrics recommends recalibration of the aluminum oxide moisture
sensor on the Replaceable Transducer Element (RTE) every 6 to 12 months. The optimum interval depends on the
specific application. To accomplish this, either return the RTE to Panametrics for recalibration or install a new RTE. The
HygroPro
II
electronics will automatically read and store the calibration data whenever a new or recalibrated RTE is
installed.
Error Code
Description
On screen Message
E1
Error happens when probe can't be found
No Probe Error
E2
When measurement is out of the defined calibration curve range
Out of range Error
E3
When measurement is above the defined calibration curve range
Over Range Error
E4
When measurement is below the defined calibration curve range
Under Range Error
E7
This error happens when the link to the probe is dead
No Link Error
E8
The message Probe received does not have the correct CRC
Bad CRC Error
E13
The calibration curves were never entered, they have their zero
default entries
No Calibration
E19
This error happens when there is a hardware fault with the moisture
sensor
Hygro Sensor Fault
E20
This error happens when there is a hardware fault with the
temperature sensor
Temp Sensor Fault
E21
This error happens when there is a hardware fault with the pressure
sensor
Pressure Sensor Fault
E26
The ADC for the pressure measurement failed
ADC Failure
CAUTION!
Be sure to perform the probe cleaning procedure in a well-ventilated area or laminar
fume hood. Observe all necessary safety precautions when handling the cleaning
solvents.
Summary of Contents for HydroProII
Page 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Preface x HygroProII User s Manual no content intended for this page...
Page 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Page 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Page 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Page 103: ......