Chapter 1. Installation
8
HygroPro
II
User's Manual
1.4.1.3
Standard Connections with a PC for field software upgrade
Refer to Figure 9 on the next page and Table 2 below, and complete the following steps to wire the transmitter.
1.
Connect an RS485-to-USB adapter (customer-supplied) to an available serial port on the PC.
2.
Push the female connector end of the factory-supplied cable into the mating male connector on the transmitter
module. Make sure the pins are properly aligned. Then, secure the connectors together by sliding the metal
sleeve on the cable over the connectors and turning it clockwise until it is tight.
3.
Using the flying leads at the other end of the factory-supplied cable, connect the transmitter to the
RS-485-to-USB adapter. The cable shield should be connected to the common terminal.
4.
Trim any unused leads even with the outer cable jacket to remove any bare wire and prevent accidental short
circuits.
IMPORTANT:
To install the HygroPro
II
in a hazardous (classified) area,
see “Hazardous Area Wiring Connections” on
Figure 9: Standard connections with a PC
Table 2: Cable Leads - With a PC
Lead
Connection Description
Blue
Power Supply (+) [12-28 VDC]
Brown
Power Supply (–) [Return]
White
RS485-RS232 Converter (+) [positive]
Black
RS485-RS232 Converter (–) [negative]
Ground
Earth Ground
Summary of Contents for HydroProII
Page 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Preface x HygroProII User s Manual no content intended for this page...
Page 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Page 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Page 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Page 103: ......