HygroPro
II
User's Manual
55
Appendix A. HygroPro
II
HART
®
Field Device Specification
Command-Specific Response Codes
A.9.10 Command 149 (0x95): Read Temperature Constant
Reads the temperature constant used instead of the live measurement.
First byte is an enumeration: 0 = temperature constant active, 1 = live measurement instead of constant
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
A.9.11 Command 150 (0x96): Set Dew Point Offset
This command will set the Dew Point Offset adjustment.
Request Data Bytes
Code
Class
Description
0
Success
No Command-Specific Errors
1-4
Undefined
5
Error
Too Few Data Bytes Received
6
Error
Device-Specific Command Error
7
Error
In Write Protect Mode
8-15
Undefined
16
Error
Access Restricted
8-127
Undefined
Byte
Format
Description
None
Byte
Format
Description
0
Unsigned-8
0 = use constant 1 = live
1 ~ 4
Float
Temperature constant used instead of live measurement.
Code
Class
Description
0
Success
No Command-Specific Errors
1-127
Undefined
Byte
Format
Description
0
Float
Dew Point offset adjustment.
Summary of Contents for HydroProII
Page 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Preface x HygroProII User s Manual no content intended for this page...
Page 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Page 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Page 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Page 103: ......