Chapter 2. Operation
24
HygroPro
II
User's Manual
If Special is chosen at the previous prompt, the
screen shown on the left appears.
Select Zero and press Enter. Use the arrow keys to
enter the zero value for the special output and press
Enter.
Repeat the above two steps to enter the Span value
for the special output.
In the Output menu, select Range and press Enter.
The screen shown on the left appears.
Enter the Zero and Span values for the range, using
the same procedure as above.
In the Output menu, select Trim and press Enter. The
following screen appears:
Enter your values for the Base Trim, Span Trim and %,
using the same procedures as on the previous
screen.
When you select the Mode option at the above
prompt, choose either Test to verify the output or
Normal for normal operation.
In the Test mode, you can check (and trim if
necessary) the output values at any desired % level
between 0 and 100%.