Appendix A. HygroPro
II
HART
®
Field Device Specification
52
HygroPro
II
User's Manual
A.9.5 Command 140 (0x8C): Read Primary Variable Range Values
This command is to read the PV range.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
A.9.6 Command 144 (0x90): Set Loop Measurement Type
This command is to set the loop measurement type to 1 of 3 enumerations:
0 – 4-20 mA, 1 – NAMUR, 2 – Special Zero and Span Values
When the loop measurement type is set to Special, then the Special span value and special zero value will determine
the endpoint mA values instead of 4 mA and 20 mA.
Request Data Bytes
Response Data Bytes
32
Error
Busy
33 ~ 127
Undefined
Byte
Format
Description
None
Byte
Format
Description
0
Unsigned-8
Upper and Lower Range Values Unit Class
1
Unsigned-8
Upper and Lower Range Values Unit Code
2 ~ 5
Float
Upper Range Value
6 ~ 9
Float
Lower Range Value
Code
Class
Description
0
Success
No Command-Specific Errors
1 ~ 127
Undefined
Byte
Format
Description
0
Unsigned-8
Loop Measurement type
1~4
Float
Special Zero Value
5~8
Float
Special Span Value
Byte
Format
Description
0
Unsigned-8
Loop Measurement type
Code
Class
Description
Summary of Contents for HydroProII
Page 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Preface x HygroProII User s Manual no content intended for this page...
Page 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Page 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Page 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Page 103: ......