Appendix A. HygroPro
II
HART
®
Field Device Specification
58
HygroPro
II
User's Manual
Response Data Bytes
Command-Specific Response Codes
A.9.16 Command 155 (0x9B): Write Hygro Calibration Curve
This command is to read the loop 0 current output error handling.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0
Unsigned-8
Index of Hygro curve point.
1 ~ 2
Signed-16
Dew Point Value of point.
3 ~ 6
Float
FH value of point.
Code
Class
Description
0
Success
No Command-Specific Errors
1
Error
Index Out of Bounds.
2 ~ 127
Undefined
Byte
Format
Description
0
Unsigned-8
Index of Hygro curve point.
1 ~ 2
Signed-16
Dew Point Value of point.
3 ~ 6
Float
FH value of point.
Byte
Format
Description
0
Unsigned-8
Index of Hygro curve point.
1 ~ 2
Signed-16
Dew Point Value of point.
3 ~ 6
Float
FH value of point.
Code
Class
Description
0
Success
No Command-Specific Errors
1-3
Undefined
4
Error
Index Out of Bounds Error.
5
Error
Too Few Data Bytes Received
6
Error
Device-Specific Command Error
7
Error
In Write Protect Mode
8-15
Undefined
16
Error
Access Restricted
8-127
Undefined
Summary of Contents for HydroProII
Page 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Preface x HygroProII User s Manual no content intended for this page...
Page 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Page 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Page 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Page 103: ......