3. Troubleshooting
Problem
Cause
>>
possible solution
●
WEISSCAM HS-2 does not start
No power source connected
>> Connect power source
Power source must supply the correct
voltage (12 - 24V DC), above all for bat-
tery operation
>> Replace battery
Cable defective or incorrect pin assign-
ment
>> Check / replace cable
Turn-on pulse not received correctly
>> Turn the WEISSCAM HS-2 off and back
on again
●
●
●
No image or faulty image at the
HD-SDI LINK A / LINK B
Debayerboard started incorrectly
>> Turn the WEISSCAM HS-2 off and
back on again
Respective port defective
>> Plug the BNC cable into the second
HD-SDI port (single link)
BNC cable defective
>> Replace BNC cable
Target device defective or incompatible
>> Connect to another target device
(e.g. monitor)
Signal transmission on the BNC cable
disrupted by another cable
>> Check the cable routing and bypass
possible sources of interference
●
●
●
●
●