Select “Factor” and enter the “Modify” screen.
Adjust the factor until the power/energy reading on
the screen equals the known power/energy. Press
“Save” and the power/energy calibration factor
under “Updated” will change accordingly. See Note
1.
For units which have different calibration factors, e.g.
CO2, YAG, or VIS, press "Laser" to select correct type
of laser before calibration. When recalibrating using
"Specific: this laser only“ function, only the
calibration of the chosen laser wavelength will
change. If you wish to change the calibration of all
wavelengths proportionately, then use the function
"Overall: all lasers ".
Note 2:
When changing power calibration, this will also
change energy calibration proportionately. However,
changing energy calibration will not change power
calibration.
Note 1: