Detailed information about the system
10
97
System Accuracy Results for Glucose Concentrations >4.2 mmol/L
Percent (and number) of meter results that match the laboratory test
Within ±5%
Within ±10%
Within ±15%
Within ±20%
64.2% (308/480)
91.5% (439/480)
99.0% (475/480)
99.8% (479/480)
System Accuracy Results across the entire Glucose Range
Percent (and number) of meter results that match the laboratory test
Within ±0.83 mmol/L or ±20%
99.7% (598/600)
Therefore, 99.7% of the total results obtained with the OneTouch® Verio™Pro+ System achieved the goal suggested
by the diabetes experts.
Yellow = 5mm margin, Red = 7mm margin
Yellow = 5mm margin, Red = 7mm margin
AW05948101B_OT_V_OB_Insides_R1.indd 97