Troubleshooting
9
86
(continued)
What it means for Linearity Solution Testing
Meter is too cold (below 18°C) to work correctly.
What to do for Linearity Solution Testing
Move the meter and test strips to a warmer area. Insert a new test strip when the
meter and test strips are within the operating temperature range (18°C-28°C).
If you do not get another “Temperature too low” message, you can proceed with
testing. If the error message continues to appear 30 minutes after moving to a
warmer area, contact Customer Service and/or LifeScan® sales representative.
Warning
Temperature
too low. Out of
operating range.
See Meter
Operator’s Guide.