Troubleshooting
9
92
What it means
The meter has detected a problem with the test strip. Possible cause is test strip
damage.
What to do
Repeat the test with a new test strip; see blood application (pages 50-54), control
solution testing (pages 22-31), or linearity testing (pages 62-73). If the error message
appears again, contact Customer Service and/or LifeScan® sales representative.
ERROR 5
Strip problem.
Retest with a
new strip.
What it means
Meter battery is low but still has enough power to perform a test.
What to do
Once the
Low Battery
message is displayed, this symbol will continue to appear
until you replace the batteries. Test results will still be accurate, but replace the
batteries as soon as possible; see pages 80-82.