Troubleshooting
9
84
Warning
Temperature
too high. Out of
operating range.
See Meter
Operator’s Guide.
(continued)
What it means for Linearity Solution Testing
Meter is too hot (above 28°C) to work correctly.
What to do for Linearity Solution Testing
Move the meter and test strips to a cooler area. Insert a new test strip when the
meter and test strips are within the operating temperature range (18°C-28°C). If
you do not get another “Temperature too high” message, you can proceed with
testing. If the error message continues to appear 30 minutes after moving to a
cooler area, contact Customer Service and/or LifeScan® sales representative.