24
Maintenance
Maintenance Inspection and Safety Management
The HBP-1300 must be maintained to ensure functionality and to secure the safety of patients and operators.
Daily checks and maintenance should be performed by the operator. (page 25)
In addition, qualified personnel are necessary to maintain the performance and the safety, and to conduct
periodic inspections. We recommend that the verification test be performed at least once a year.
Cleaning of the Device
Cleaning and disinfecting should be performed in accordance with your facility’s infection control
practice.
Surface cleaning
Wipe with a cloth that has been moistened with isopropyl alcohol diluted to 50 v/v%, or ethyl alcohol
(disinfection alcohol) diluted to 80 v/v% or less and wrung out.
Do not wipe the DC jack or allow it to become wet.
Removing dust
Use a moistened cotton bud to remove dust that has accumulated on the vent ports.
Service
The device requires no routine service other than cleaning, and visually checking the cuffs, tubing, etc.
Caution
• Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-concentration ozone, etc.).
• If using an antiseptic solution for cleaning, follow the instructions of the manufacturer.
Accessory Care
Non-Invasive Blood Pressure (NIBP) Measurement
Cuff
Wipe clean on the surface of the cuff with a cloth moistened with a 70 v/v% dilution of isopropyl alcohol,
or a 80 v/v% or less dilution of disinfection ethanol (ethyl alcohol).
Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the inside well.
Example of connection for pressure accuracy confirmation:
1. Display the pressure accuracy confirmation screen as explained in
“Utility Mode” on page 17.
2. Connect the blood pressure monitor, the calibrated reference pressure
gauge, and the cuff and inflation bulb.
3. Check the pressure value of the blood pressure monitor and the
pressure value of the calibrated reference pressure gauge.
Note:
• Make sure that the blood pressure monitor reading is within ±3 mmHg compared to the calibrated
reference pressure gauge.
• To rapidly deflate the cuff, press the [ ] button.
To repeat “Pressure Accuracy Confirmation”, turn off the power and repeat the procedure from
step 1 in “Utility Mode” on page 17.
Display the pressure value.
Display “0” in the pressure accuracy confirmation screen.
HBP-1300-E main.book Page 24 Thursday, February 11, 2016 11:39 AM
Summary of Contents for HBP-1300
Page 232: ...230 HBP 1300 E main book Page 230 Thursday February 11 2016 11 39 AM...
Page 233: ...231 RU OMRON HBP 1300 E main book Page 231 Thursday February 11 2016 11 39 AM...
Page 234: ...232 IEC 60601 1 2 OMRON HBP 1300 E main book Page 232 Thursday February 11 2016 11 39 AM...
Page 235: ...233 RU SpO2 HBP 1300 E main book Page 233 Thursday February 11 2016 11 39 AM...
Page 236: ...234 HBP 1300 E main book Page 234 Thursday February 11 2016 11 39 AM...
Page 237: ...235 RU 6 20 30 C 4 86 F 65 20 HBP 1300 E main book Page 235 Thursday February 11 2016 11 39 AM...
Page 239: ...237 RU 241 252 OMRON 5 HBP 1300 E main book Page 237 Thursday February 11 2016 11 39 AM...
Page 244: ...242 1 2 3 4 5 4 300 4 HBP 1300 E main book Page 242 Thursday February 11 2016 11 39 AM...
Page 253: ...251 RU START STOP 3 HBP 1300 E main book Page 251 Thursday February 11 2016 11 39 AM...
Page 257: ...255 RU 30 HBP 1300 E main book Page 255 Thursday February 11 2016 11 39 AM...
Page 264: ...262 HBP 1300 E main book Page 262 Thursday February 11 2016 11 39 AM...
Page 308: ...HBP 1300 HBP 1300 HBP 1300 HBP 1300 d 1 2 P d 1 2 P d 2 3 P d P...
Page 311: ...AR EMC EMC EN60601 1 2 2007 EN60601 1 2 2007 OMRON HEALTHCARE EN60601 1 2 2007...
Page 312: ...NIBP EN1060 1 1995 A2 2009 EN1060 3 1997 A2 2009 ISO81060 1 2007 K5 93 42 EEC EN 1060...
Page 314: ...E3 OMRON E40 E41 E42 OMRON ABS...
Page 315: ...AR E40 E2 E9 OMRON NIBP E1 OMRON GS CUFF E2...
Page 316: ......
Page 317: ...AR OMRON OMRON OMRON START STOP NIBP...
Page 318: ...START STOP NIBP OMRON GS CUFF...
Page 319: ...AR HBP 1300 80 v v 50 v v NIBP 70 v v 80 v v...
Page 320: ...START STOP...
Page 321: ...AR ON MODE MODE...
Page 322: ...START STOP...
Page 323: ...AR INDEX ARTERY RANGE INDEX ARTERY RANGE INDEX ARTERY...
Page 324: ...GS CUFF XL GS CUFF L GS CUFF M GS CUFF S GS CUFF SS NIBP OMRON GS CUFF CUFF HOSE CUFF HOSE...
Page 326: ...START STOP MODE E40 MODE START STOP...
Page 327: ...AR MODE 280 AUTO 220 250 280 220 AUTO 280 250 220 AUTO MODE ON MODE START STOP OFF AUTO ON OFF...
Page 328: ...LCD LCD E40 E40 OMRON...
Page 329: ...AR...
Page 330: ...LCD...
Page 331: ...AR START STOP MODE NIBP BF...
Page 333: ...AR OMRON M GS CUFF L HXA BAT 2000 9065797 0...
Page 334: ...OMRON NIBP...
Page 335: ...AR NIBP luer IBP SpO2 OMRON GS CUFF...
Page 336: ......
Page 337: ...AR...
Page 338: ...SpO2...
Page 339: ...AR IEC60601 1 2 OMRON...
Page 340: ...OMRON MRI LCD...
Page 341: ...AR...
Page 342: ...OMRON OMRON OMRON OMRON OMRON OMRON OMRON OMRON...
Page 343: ...NIBP HBP 1300 AR...